A comparative study on calculation of intraocular lens power using different formulas between IOLMaster 700 and IOLMaster 500 in cataract eyes

Authors: Deng Xiaohui,  Chang Pingjun,  Huang Jinhai,  Wang Dandan,  Zhao Yinying,  Ding Xixia,  Zhao Yun’e
DOI: 10.3760/cma.j.cn115989-20200226-00110
Published 2022-12-10
Cite as Chin J Exp Ophthalmol, 2022, 40(12): 1170-1175.

Abstract                              [View PDF] [Read Full Text]

Objective

To compare the accuracy of IOLMaster 700 and IOLMaster 500 in intraocular lens (IOL) power calculation.

Methods

A cross-sectional study was conducted.Two hundred and sixty-two eyes of 262 patients who underwent phacoemulsification combined with IOL implantation at the Eye Hospital of Wenzhou Medical University from November 2018 to November 2019 were enrolled.Preoperative biometry for cataract surgery was performed using IOLMaster 700 and IOLMaster 500.IOL power was calculated through the built-in formulas, Haigis, Holladay Ⅰ, Hoffer Q and SRK/T of the two devices.The difference in IOL power calculation between the two devices was analyzed through the prediction error of IOL power calculation using different formulas across different axial length (AL) ranges.This study complied with the Declaration of Helsinki.The study protocol was approved by the Ethics Committee of the Eye Hospital of Wenzhou Medical University (No.2020-038-K-33). Written informed consent was obtained from each patient before the surgery.

Results

There was no significant difference in mean absolute error (MAE) between IOLMaster 700 and IOLMaster 500 using Haigis, Hoffer Q and SRK/T over the entire AL range (all at P >0.05). The MAE of IOLMaster 500 was 0.47 (0.24, 0.90) D, which was significantly lower than 0.50 (0.28, 0.99) D of IOLMaster 700 using Holladay Ⅰ formula (Z=-3.120, P=0.002). When AL was <22.0 mm and ≥24.5 mm-<26.0 mm, there was no significant difference in MAE between the two devices using the four formulas (all at P >0.05). When AL was ≥22.0 mm-24.5 mm, there was no significant difference in the MAE between the two devices using Haigis, Hoffer Q and SRK/T (all at P >0.05), but 0.42 (0.18, 0.75) D from IOLMaster 500 was smaller than 0.45 (0.25, 0.79) D from IOLMaster 700 using Holladay Ⅰ, showing a statistically significant difference (Z=-3.487, P <0.001). But the difference was negligible and therefore was of no clinical significance.When AL was ≥26.0 mm, there was no statistically significant difference in the MAE between the two devices using Haigis, Holladay Ⅰ and SRK/T, but 0.66 (0.38, 1.00) D from IOLMaster 500 was significantly smaller than 0.98 (0.62, 1.32) D from IOLMaster 700 using Hoffer Q (Z=-3.046, P=0.002).

Conclusions

The refractive prediction accuracy of IOLMaster 700 and IOLMaster 500 using Haigis, Hoffer Q and SRK/T is similar over the entire AL range.For patient with long AL, the IOL calculation from IOLMaster 700 using Hoffer Q is significantly larger than that from IOLMaster 500, which requires extra caution in clinical practice.The accuracy of IOLMaster 700 and IOLMaster 500 for IOL prediction is very similar.

Key words:

Cataract; Lenses, intraocular; Optical biometry; Refractive power; IOLMaster; IOL calculation formula

Figures & Tables

Contributor Information

Deng Xiaohui

The Eye Hospital of Wenzhou Medical University, Hangzhou 310020, China

Deng Xiaohui is now working at Ningbo No.2 Hospital, Ningbo 315010, China

Chang Pingjun

The Eye Hospital of Wenzhou Medical University, Hangzhou 310020, China

Huang Jinhai

The Eye Hospital of Wenzhou Medical University, Hangzhou 310020, China

Huang Jinhai is now working at Eye Institute and Department of Ophthalmology, Eye &

ENT Hospital, Fudan University, Shanghai 200031, China

Wang Dandan

The Eye Hospital of Wenzhou Medical University, Hangzhou 310020, China

Zhao Yinying

The Eye Hospital of Wenzhou Medical University, Hangzhou 310020, China

Ding Xixia

The Eye Hospital of Wenzhou Medical University, Hangzhou 310020, China

Zhao Yun’e

The Eye Hospital of Wenzhou Medical University, Hangzhou 310020, China

(Read 91 times, 1 visits today)