Imaging quality and anti-eccentricity performance of axial modulation aspherical intraocular lens in model eyes

Authors: Zhang Bin, Ren Taoran, Wu Jiahui, Li Minhao, Liu Danyan, Liu Xiaoli, Geng Yuxin
DOI: 10.3760/cma.j.cn115989-20240531-00139
   

Citation

Zhang Bin, Ren Taoran, Wu Jiahui, et al. Imaging quality and anti-eccentricity performance of axial modulation aspherical intraocular lens in model eyes[J]. Chin J Exp Ophthalmol, 2026, 44(3):227-231. DOI: 10.3760/cma.j.cn115989-20240531-00139.

ABSTRACT                   [Download PDF]  [Read Full Text]

Objective  To investigate the imaging quality and anti-eccentricity performance of axial modulation aspherical intraocular lens (IOL) in model eyes.

Methods  A ray-tracing program of optical simulation software was used to evaluate the imaging quality of four different IOLs, including PRS60A IOL, Tecnis Z9000 with -0.27 μm spherical aberration aspherical IOL, SofPort AO with 0 μm spherical aberration aspherical IOL and SA60AT spherical IOL under the condition of either centration or decentration. Modulation transfer functions (MTF) under pupil diameters of 3.0, 4.0, and 5.0 mm and IOL decentrations of 0, 0.25, 0.50, and 0.75 mm were calculated and plotted. A Monte Carlo analysis was performed after 1 000 simulations of IOL decentration, ranging from 0 to 0.5 mm, randomly conducted for every IOL.

Results  Under 3.0, 4.0, and 5.0 mm pupil diameters, the centered Tecnis Z9000 IOL had the highest MTF value and the best imaging quality, and when the eccentricity was greater than 0.50 mm, the MTF value decreased significantly, and the imaging quality was significantly reduced. In most cases, the PRS60A IOL was superior to the SofPort AO and SA60AT IOLs when centered and eccentric at various pupil diameters. Monte Carlo analysis showed that the postoperative effects of PRS60A IOL were relatively repeatable and predictable, and the overall imaging quality was better than that of a fixed spherical aberration IOL.

Conclusions  Compared with the fixed spherical aberration IOL, the progressive modulation type aspheric IOL of axial spherical aberration has better imaging quality and tolerance to eccentricity errors in model eyes.

Cataract; Intraocular lens; Decentration; Modulation transfer function; Optical performance; Modelling

Authors Info & Affiliations 

Zhang Bin
Department of Ophthalmology, the Second Hospital of Hebei Medical University, Shijiazhuang 050000, China
Ren Taoran
Department of Ophthalmology, the Second Hospital of Hebei Medical University, Shijiazhuang 050000, China
Wu Jiahui
Department of Ophthalmology, the Second Hospital of Hebei Medical University, Shijiazhuang 050000, China
Li Minhao
Department of Ophthalmology, the Second Hospital of Hebei Medical University, Shijiazhuang 050000, China
Liu Danyan
Department of Ophthalmology, the Second Hospital of Hebei Medical University, Shijiazhuang 050000, China
Liu Xiaoli
Department of Ophthalmology, the Second Hospital of Hebei Medical University, Shijiazhuang 050000, China
Geng Yuxin
Department of Ophthalmology, the Second Hospital of Hebei Medical University, Shijiazhuang 050000, China
(Read 1 times, 1 visits today)